More detector, dewar and Arcon work - Ramon Galvez , August 16th, 2002
Done:
-
Tested the new macro thresholds set for cryo alarms. Working ok.
-
With Rolo C. we made changes in the Hydra option of ‘SetFiles’. The waveforms and sequencer code are automatically installed now when Hydra + Arcon3.7 are selected.
-
Tests were done also when switching back and forth for other ccds.
-
With Ricardo V., we smoothly went through the LN2 solidification using the new device:
Normal operation vac sensors readouts
CCD dwr vac before pumping |
: 300 Torr |
CCD dwr vac after pumping |
: 0 mT |
|
|
LN2 flask vac after refill |
: 400 Torr |
LN2 flask vac for solid state |
: 20 Torr ( after 1hr pumping ) |
To be done after this run (minor pending details) :
-
Improve the ‘solid N2’ valve setpoint lock device.
-
Install the new adjustable security ‘pop-up’ pressure valve.