Revised 13 July 2020 by A. R. Walker.
Original by Maxime Boccas, 18 October 1999.
Over the period 1993-2000, many improvements have been carried out at the Víctor M. Blanco 4-meter Telescope to improve its image quality. The principal changes are: an active primary mirror support with lookup tables, a refigured f/8 secondary mirror, an image analyzer for the Cassegrain foci, removal of most heat sources inside the dome, control of the dome temoerature during the day using the AHUs and heicopter fan, active ventilation (air sucker) of the primary cell at night, dome ventilation doors and active control of oil temperature. The dome outer surface is covered with insulated aluminum panels which although the outer surface heats up during the day mpre than the originbal Lomit paint, does not over-cool at night..
Since 2012, DECam at Prime Focus has introduced a 5-axis hexapod (x,y,z, tip, tilt) position of the instrument and corrector, driven by a Look Up Table and tweaked on an exposure by exposure basis by a wavefront sensor that analyses out-of-focus inages, and makes a "tweak" to the hexapod position.
The night assistant must check that all the environmental control functions explained below are working properly during the night.
1.1 General closure conditions:
ABSOLUTE closure when:
Dome re-opening:
GUIDELINE :
BEWARE PLEASE: Don't permit observing beyond these limits! If you have trouble to be heard by the astronomer, please REPORT it (so we can politely explain the rules).
The dome AND the mirror cover have to be opened 1 hour before sunset, orienting the shutter eastward to make sure no direct sunlight will enter the dome. Dome and mirror cover have to be closed at the end of the night.
1.3. Lateral sliding doors opening:
These doors have to be opened at sunset and left opened all night.
The Primary mirror cooling system has from 2019 been activated only in "sucking" mode, all the time, day and night. That is,we do no longer blow cold air onto the primary.
During the night, you are invited to check Blanco ECS Telemetry [3] and check that:
1.5. Floor cooling, Dome Air Handling Units (AHUs), and stirring fan on main floor (M):
The cooling of the C floor and M floor in the dome is always ON. It is currently regulated automatically.
The two AHUs should always be off when the dome is open. There is no automatic control, Thus the procedure is for the telescope operator to turn green on at the end of the night, and then Telops staff evaluate the inside versus outside temperature and the weather forecast prediction for the following night, then decide whether to turn off one or both, and at what time.
BEWARE: In conditions of high humidity the AHUs can ice up and restrict the air flow, This condition can be cleared rapidly by turning off the cooling, air flow on.
The large stirring fan (helicopter) on the M floor (inside the dome) should always be ON when the dome is closed and OFF when the dome is opened. This fan improves greatly the air circulation inside the dome.
SAFETY: Both the AHUs and the helicopter can be turned off during instrument changes or any other works around the telescope if the noise is a safety concern, e.g. by impeding communication. But don't forget to turn them back on when you are done!
The cooling of the oil circulating under the R.A. rear and front pads is activated automatically when the pumps are turned ON by Observer support at sunset. The oil glycol valve is controlled (range is 1-2.25 V) by an equation involving the low dome and oil glycol temperature sensors. It is important that the oil cooling be working: if it is not, you can diagnostic it in ECS Telemetry [3] (the "before pad" and "glycol" entries on the OIL line will quickly indicate around 30°).
The oil temperature for the Hidrostatic Bearings must be between 8 to 10C.
1.7. Chimney fans: Not in use.
These fans are located inside the chimney at the level of the primary mirror and suck ambient air into the mirror cell so as to form a laminar downward flow in the chimney. This is to prevent Cass cage heat sources from creating convection in the light path in the chimney.
At Cassegrain foci (f/8), Observer Support has to turn ON the chimney fans before sunset. The switch is on the power supply on the old console room desk. The night assistant must turn it OFF at the end of the night.
At Prime Focus: The top of the chimnney is covered with a cap.
The current thermal plan in the building is not changed. Always leave the air conditioning at full power. Do not use heat sources at level M, MZ and C. Always maintain the doors closed in the passageways (especially the glass door in front of the lift at GR floor, the cryocooler compressor and pump rooms). In general, respect the signs in the building.
2. Brief reminder of known optical problems
The image quality of the telescope can suffer deterioration mainly from 3 optical problems:
Note: IMAN is normally used only on engineering nights.
This is due to tiny slippage of M1 and/or deformation under its own gravity. Astigmatism will show up as elongated images, perpendicular on each side of the focus, especially for large H.A. (typically more than 2 hours). We compensate that with a lookup table, which controls the pressure in the air bags under M1. Thus the TCP toggle "CORR" should always be ON, at all foci. Note that if the F/8 secondary mirror lose its vacuum, strong astigmatism will immediately show up (typically 2-4 microns as measured by iman).
This aberration shows up in 2 cases: at Prime Focus if the primary mirror axis is not aligned with the instrument and in the 2-mirror configuration (f/8) when there is a misalignment between them. Coma produces images with a flare (a tail like a comet) which is identical (same amplitude and direction) on each side of the focus.
At Prime Focus coma is controlled by the DECan active optics system and Look Up Table that control the hexapod on which the instrument is mounted.
Coma usually increases the further you go from the optical axis -this is called "field coma" and is normal (it is always supposed to be 0 on axis)- and also shows up when the optical axis of M1 and M2 are not coaligned -this is called "decentering coma" and is abnormal-. Our f/8 is a Ritchey-Chretien (RC) type, which means it is optically designed to correct the field coma. If the telescope is properly aligned there should not be any visible coma at f/8. If coma shows up anywhere in the field (it would usually be constant all over the field), there is a mirror misalignment and this can be corrected by using IMAN.
Thermal expansion that can occur at night will cause the telescope to defocus. The prime focus changes at a rate of -110 microns/°C, the f/8 focus at -780microns/°C. The actual movement of He f/8 mirror is a factor 10 less (i.e. -78 microns/°C). The auto[focus of DECam means you do not have to worry about focus at prime focus.
Use Blanco ECS Sensors Diagram [1] to check the Serrurier truss temperature and adjust the focus accordingly (F/8).
3. Prime Focus nights:
Focus decreases by 110 units per 1° increase in temperature.
4. f/8 nights:
* CORR ON all the time.
* M2 Corrections ON all the time
* Temperature drift compensation:
Focus number increases when focal plane (and M2) moves up
The focus number represents microns of motion of the focal plane (not of M2)
Decrease focus setting 780 units per 1° increase in temperature
*Collimation:this has to be checked according to the following procedure:
Each material has its own radiation cooling property (which is not easy to calculate by the way). Just know that if there is no wind, the radiation cooling is low and surfaces will not equilibrate thermally with the ambient temperature very quick. Also, a surface, even with moderate ability to radiate, looking into a "cold" sky can cool down several degrees below ambient temperature. Fortunately, a mirror radiates very little. Nevertheless, when cooling the mirror during the day, the mirror temperature is usually several degrees below ambient temperature and there are risk of condensation. The control loop will turn off the cooling when the difference between the mirror temperature and the dew point shrinks to only 2 degrees.
Knowing the Relative Humidity (RH in %) and the ambient temperature (Ta) of the air, one can calculate the dew point temperature (Td) and determine whether a surface at temperature Ts can become wet or not :
I think "dew point" refers only to the case where RH=100%; when RH<100% we might have to talk about condensation point. We won't do that distinction anymore in the text.
How to use the table?
From your ambient temperature on the X-axis, go up to intersect the curve corresponding to the RH, then move horizontally onto the left to read the dew point on the Y-axis. You can then check if your surface is hotter or colder than this dew point value.
Example: Ta is 10°C, RH is 80%, Td is 6.5°C. So a 5°C mirror would be wet!
Calculation of dew point:
PS = (5.10-7.T3 + 10-5.T2 + 5.10-4.T + 0.0061) / 1.294
Td = 6.108.PV5 - 108.PV4 + 6.106.PV3 - 193789.PV2 + 3957.9.PV -14.911
The nightly procedures are carryed out by the following TelOps teams: Observer Support, Telescope Operators and Floater Operators.
Floater operators can performed the Observer Support or the Telescope Operators tasks.
There are 3 procedures well defined in order to operate efficiently and safe:
Only for internal use, the documents linked in 1 and 3 are from the TelOps NAS storage disk. If you have access go to 139.229.13.92/TOLNAS0/TelOpsDocs/Procedimientos/Observer Operators.
Links
[1] http://ctio4l.ctio.noao.edu/web/Blanco/diagram.html
[2] http://www.ctio.noao.edu/noao/content/Relative-Humidity
[3] http://ctio4l.ctio.noao.edu/web/Blanco/graph.php
[4] http://ctio4l.ctio.noao.edu/web/Blanco/ecs/Oil_Cooling_schema.png
[5] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/dewpoint.gif
[6] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/ctio/4m/Blanco_Opening_checklist_V2.pdf
[7] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/ctio/4m/Checklist4m_endofnight3.pdf
[8] http://www.ctio.noao.edu/noao/content/Thermal-Control-Procedures