CTIO
Published on CTIO (http://www.ctio.noao.edu/noao)

CTIO Home > Optical Engineering > Maintenance Procedures > Aluminizarion & Washing

Aluminizarion & Washing

FILES

4.0-m Aluminizing Procedure

This operation is normally conducted every 2 years in average during the winter shut-down of August.

Blanco Aluminization test 2009 summary [1]

The entire process takes about 6 days (from the moment the mechanics team start until they leave the telescope ready for optical alignment). This is a picture [2] of the 1998 aluminizing team.

Before the aluminizing chamber (start with the mirror set up on its pedestal in the washing area)

Inside the aluminizing chamber

 

Before the aluminizing chamber  (4 hours typ.)

This is done ideally with 6 people: 3 outside the mirror, 2 inside the central hole, 1 preparing materials.

Here is a list of materials [3]. THINK SAFETY! Wear plastic suit all the time. Wear heavy gloves, goggle and respiratory mask when manipulating acids.

Have a plan for what to do in case of acid accident.

 

 

A. MIRROR PREPARATION AND WASHING

A.0. Pour natural sponges in a bath of HCl+H2O (to disolve any residuals of shells), rinse with water and cycle 5 times in washing machine.

A.1. For rough degrease: 200 gr of soap into 10 liters of filtered water (poured into 3 buckets)

A.2. For removing aluminium : 10 liters HCl + 10 liters filtered water + 200 grams CuSO4. !!CHANGE: a 15% HCl solution is enough (see how to apply it below)!!

A.3. For final degrease: 1 kg KOH into 20 liters of filtered water

A.4. Clean the mirror with CO2 snow, measure reflectivity and photograph mirror

A.5. If necessary, lay lab wipes on oil spots to absorb oil

A.6. Wear surgical gloves

A.7. From now onto the drying phase, ALWAYS keep the mirror wet.

Rinse entire mirror with filtered water (3 hoses) for 5 min. NO contact.

A.8. Pour soap onto entire mirror around the circumference. NO contact. Rinse with filtered water. REPEAT this step twice.

A.9. Blot mirror with soapy sponges using only the weight of the sponge. NO wiping. 3 people on outside of mirror make 2 complete circles. 2 people in the central hole make 3 complete circles.

A.10. Wash sides of mirror with soap and dedicated sponge NOT to be used on mirror optical surface.

A.11. Rinse the entire mirror with filtered water.

A.12. REPEAT step A.9 to A11.

alum10.jpg [4]

Figure 1: Mirror preparation & washing

 

B ALUMINIUM REMOVAL AND DEGREASING

B.1. Wear new surgical gloves + heavy gloves + goggle + respirator

B.2. One person pours HCl/CuSO4 around circumference of mirror while the 5 others drag-wipe with balls of cotton. Each cotton ball is used only for a maximum of 3 drag-wipe actions then replaced. Continue until aluminium is removed. !!CHANGE: cover the mirror with Kimwipes and pour HCl solution on them, they will keep the acid longer in contact with the coating and dissolve it without the need for a large quantity of acid!! Residual Al spot can we drag-wiped individually with a Kimwipe ball.

B3. Rinse with filtered water until pH is neutral. Rinse 2 minutes more with filtered water.

B4. Change to clean surgical gloves

B.5. One person pours KOH around circumference of mirror while the 5 others drag-wipe with cotton balls. Each cotton ball is used only for a maximum of 3 drag-wipe actions then replaced.

B.6. Rinse with filtered water until pH is neutral.

B.7. REPEAT steps B.5 to B.6

B.8. Rinse with filtered water 2 minutes more. Rinse carefully the mirror edge and the radial support mounts and holes.


C. DRYING (this is the tricky part !)

C.1. Change to clean surgical gloves AND dust mask

C.2. Rinse with 50 liters DOUBLE distilled water pouring around the circumference. Check out that water flows uniformly on the surface, not leaving 'holes'that would be signs of broken surface tension of the liquid passing on a still-contaminated area.

C.3. 6 people using nozzle guns blow dry nitrogen from the outer diameter to the inner, making sure the water drops flows uniformly radially inward (look at them against the lights on the other side of the mirror. Dry as fast as possible to avoid water drying on its own. No contact with mirror during that phase. Bidistilled water is clean enough that the process doesn't leave water marks. Perform the breath test to detect any drying or contamination spots in any suspicious areas (breath gently on the mirror -dry your mouth first!- and watch out any pattern on the surface covered with moisture).

C.3. BIS. Old procedure: Have at least large 50 balls of Kimwipes ready and someone preparing more if necessary. They must be thick and wide, bigger than your hand. Rub (for the first time in the entire process) the mirror applying about 1-2 kg force on the Kimwipes balls. Do not overdo it! M  he ball slowly to suck the water. Use each ball ONLY FOR ONE drag-wipe action then replace it (this is very important!!!). Work moving regularly around the mirror. As the mirror dries, you will have to work faster. Constantly check the entire mirror surface for areas that need attention. MAKE SURE your (dirty) sleeves are not touching the glass when you dry reaching far with your arm!! DO NOT let the mirror dry on its own! DO NOT wipe dry areas! Use Balzers virgin optical cotton cloth for touch ups. Check any visible stains and streaks with bright light and different viewing angles. It is hard to make the mirror better but easy to make it worse.

C4. Dry sides of the mirror and holes of the radial support mounts with Kimwipes. Use dry nitrogen to CAREFULLY blow out holes. Take care not to spray water on the optical surface.

C5. Hang up mirror in its hook. Clean with acetone the bottom of the mirror especially the 3 pads. Dry bottom of mirror with Kimwipes. DO A THOROUGH cleaning of all the non optical surfaces of the mirror for complete degreasing and drying.

C.6. Setup mirror on aluminizing tank floor. Blow off dust with CO2: MAKE ABSOLUTELY sure the substrate is free of dust (other wise the coating will be full of pinholes). Put 4 clean microscope test plates (coating witness samples) on edge of mirror. Seal tank.

 

Inside the aluminizing chamber  (4 hours typ.)

Note: units used are : 1.3x10-6 atm = 1 micron (Hg) = 1 millitorr (=0.13 Pa)

Vacuum gauge positions: (to be checked)

1 is chamber

2 is holding pump

3 and 4 are the diffusion pumps

5 is roughing pump

Be careful: do not disconnect a vacuum sensor turned ON. Let any pump run for 1 min (listen the noise) before turning ON any vacuum gauge.

We indicate a typical elapsed time.

 

A. PREPARATION OF CHAMBER (a few days before)

See pictures of the inside of the chamber: (Figure 2) general view of the chamber ceiling where you can recognize : the filaments in the round holes cut in the baffle sheet, the upper window used for checking the evaporation rate of aluminium and the cable feed-though port; (Figure 3) close view showing: the 4 concentric filament annuli (number of filaments per ring, from outermost to innermost: 36, 32, 20, 16 = 104 in total), the glow discharge ring (between annulus 2 and 3), part of the entrance conduit of one diffusion pump, the nitrogen pipe (near the bottom).

 

inchamb2.jpg [5]

Figure 2: Inside the chamber

inchamb.jpg [6]

Figure 3: Inside the chamber
(close view)

 

 A.1. Work inside the chamber with clean white suit, gloves and shoes protectors. Wash the internal walls, windows and chamber floor with acetone (DON'T remove the side windows, it is hard to re-seal them). Chase any greasy spots!

A.2. Wash with HCl (strip off the aluminium) the feed-through cables contacts, the holding ring insulator.

glow.jpg [7]

Figure 4

 

A.3. For better thickness uniformity, it is recommended to put filaments only in the outer rings. ±9% uniformity is reached that way (versus ±27% with all the arrays!). Check carefully all the filaments and replace the damaged ones: knock them with the finger to see if they don't break, they must be straight when loose (i.e. unstressed), not covered with blobs of aluminium. The filaments shall not touch the baffle mesh (to avoid short circuits) and be all at the same distance above the baffle (0.25 to 0.5"). IMPORTANT: tighten the filaments with a torque wrench to about 15 in-lbs so that the current that flows in them is uniform.

A.4. Hang 3 aluminium clips at both extremities of each filaments

filament.jpg [8]

Figure 5: Filaments

 A.5. Clean and lubricate the chamber Orings with high-vacuum grease. Don't lubricate the entire flange (but just the Oring). Don't put grease in excess! Close the chamber and put the C-clamps around the perimetral flange.

A.6. Run a typical vacuum sequence as described in B.1 to B.44 with 30 minutes minimum of glow discharge and eventually filament heating -but NOT LOADED with Aluminium clips!- to outgass them all (especially the new ones).

 

B. ALUMINIZING (see picture of control panel for location of all knobs and valves, (figure 4))

4mpanelvac.jpg [9]

Figure 6: Control Panel

 

Most useful schematic diagram of the valves

panel0.jpg [10]

Figure 7: Diagram of the valves


Install the mirror and the coating monitor [11] in the chamber (check that it works!). Put test plates on the edges of the mirror along the perimeter. Close the chamber and put C-clamps.

B.1. Connect flexible pipe of roughing pump to chamber (clean and lubricate Oring) and loosen the 2 adjusting rods.

vacback.jpg [12]

Figure 8

B.2. Connect 3 cylinders of liquid nitrogen (N2): one on each diffusion pump, one for the internal circuit (Meisner trap)

B.3. Open air valve in lateral wall (red flexible pipe)

B.4. Connect water lines to the coating monitor. Open water circuit (diffusor cooler): 2 valves -input/output- on pipes in backside wall ; (roughing pump is cooled with oil)

B.5. Check air pressure on diffusion pump: must read 70-75 psi (compressor is 100psi and feeds all valves)

vacback2.jpg [13]

Figure 9

B.6.1. Connect roughing pump to energy on the wall (and unplug the M floor 'helicopter' fan). Turn ON air extractor on P floor (otherwise smoke will come out the roughing pump grey pipes).

B.6.2. Check switches: "manual" (key) and "all valves closed" (black knob). TURN ON power of control panel.

B.7. Test-cycle all valves (open/closed). The holding pump must be ON for the holding valves to work. Throttle valves leds don't shine (just watch the meter). Leave them all CLOSED including the mechanical needle valve for air inlet during glow discharge.

B.8. 0min: TURN ON roughing (mechanical) pump and immediately open foreline valves 1 and 2 (this starts pumping the diffusion pumps).

B.9. When diffusion pumps are below 20 microns (35 at least), TURN ON holding pump.

B.10. When holding pump is below 20 microns, close foreline valves 1 and 2, and open holding valves 1 and 2. Check vacuum in diffusion pumps (should remain the same at 20 microns). TURN OFF roughing pump (the holding pump is now holding the vacuum in the diffusion pumps).

B.11. TURN ON diffusion pumps 1 and 2 (it won't work if the water flow is below 3 GPM or if the diffusion pump pressure is above 25 torrs). TURN ON liquid level controllers (LN2 to diffusion pumps) and open slowly the 2 nitrogen cylinders valves (20 psi) to avoid freezing the controllers. Leave them open.

B.12. When the roughing line is at atmosphere, TURN ON the roughing pump and immediately open upper and lower roughing valves if not already opened (the roughing pump is now pumping the chamber). Check no excessive heat of diffusors.

B.13. Fan of roughing pump will power off (NOISE) automatically after 25 minutes, sign that vacuum is getting better.

B.14. 45min: When chamber pressure is around 20 microns, close upper and lower roughing valves. If vacuum in chamber doesn't seem to progress below 25 microns after some time, turn on the more reliable high vacuum gauge (cold cathode).

B.15. When the roughing line reads 10 microns, open foreline valves 1 and 2 (the roughing pump is now pumping the diffusion pump as is the holding pump).

B.16. Open throttle valves 1 and 2. Open high vacuum valves 1 and 2, waiting about 1 min between the aperture of both valves in order to prevent oil backstreaming (the diffusion pumps are now pumping the chamber).

B.17. Close holding valves 1 and 2. TURN OFF holding pump.

B.18. TURN ON high vacuum gauges (if the cold cathode is dirty, it might takes a while before it starts reading). When chamber reaches 5x10-5 torrs, TURN ON nitrogen cylinder valve for Meisner trap and corresponding liquid level controller.

B.19. 1h45min: When chamber reaches 1x10-5 torrs, reverse the process to be able to do the GLOW DISCHARGE:

B.20. Turn off high vacuum gauges, close high vacuum valves 1 and 2 and throttle valves 1 and 2.

B.21. TURN ON holding pump, close foreline valves 1 and 2. Open holding valves 1 and 2 (the holding pump pumps the diffusion pump). Check pressure in diffusion pumps.

B.22. Open upper and lower roughing valves (the roughing pump pumps the chamber).

B.23. Open air inlet needle valve by 1.25 turn and use the inlet switch to set and maintain the chamber pressure at 35 microns, which is the appropriate pressure to run the glow discharge safely. Glow discharge provides outgassing of the substrate through heating, conversion of organic substances into their volatile components and desorption of films through electrons impact.

B.24. Turn on glow discharge controller. Slowly raise the current to 3 Amps in 2 minutes. Voltage will read about 3000 Volts (if the meter works!). A uniform purple color will be visible inside the chamber. If pink arcs or flashes show up (usually at 30mic), the pressure is too low, so open air inlet valve to raise the pressure back to about 35-40 microns. The ionized air will bombard the residual sticky hydrocarbons and, by transfer of kinetic energy, help to remove them from the surfaces (especially the mirror's). Monitor the discharge for 10 to 30 minutes. Reduce the current to 0 in 30 seconds. CLOSE the needle valve by hand.

glowdis.jpg [14]

Figure 10

B.25. 2h20min: close upper and lower roughing valves. When the roughing line reads 10 microns, open foreline valves 1 and 2.

B.26. Open throttle valves 1 and 2. Open high vacuum valves 1 and 2 with the same 1-min waiting precaution (the diffusion pumps are now pumping the chamber again).

B.27. Close holding valves 1 and 2. Turn off holding pump.

B.28. TURN ON high vacuum gauges. They should read 1x10-5 torrs in a few minutes.

B.29. 2h45min: cold cathodes should read around 8x10-6 torrs.

B.30. 3h30min: cold cathodes should read at least around 6x10-6 torrs, which is just enough to aluminize (the higher the vacuum beyond that limit, the better):

B.31. Turn off all kinds of fans (Pump floor for example) to avoid vibrations in the floor.

B.32. 4h40min: Install the coating monitor [11] on top of the chamber near the window and check its parameters. With the 4 filament arrays, the chamber deposits an uneven film: 65% at r=20", 100% at r=45" and 50% at r=80". The tickness monitor is measuring at r=20", so in order to obtain 1000 Angstroms maximum thickness on the mirror, the firing should be stopped when the thickness monitor indicates 650 Ang. After you power off the filaments, the evaporation still goes on for about 70 Ang., so you should power off at about 580 Ang.

  • Using remote hand paddle from the top of the chamber, depress the "Decrease" button for at least 15sec.
  • Depress the "ON" button on the control panel to energize the outer array and zero the time counter of the Thickness monitor.
  • During the first minute, JOG the "Low Heat" button (1-2sec) of the hand paddle for small increases in power to filament up to 300A on #1. Watch filament color (yellow) through the top window and check for uniformity of all filaments.
  • JOG a few times the "High Heat" button for small increase (1sec) to reach 400A (#1) at 2'00.
  • Keep jogging "High Heat" to reach 500A at 3'00. Kepp watching the color uniformity.
  • Around 530A (3'15), the aluminium clips will start to melt and wet the filaments which will darken suddenly. The coating monitor starts showing activity: deposition has started.
  • Keep jogging "High Heat" to increase the current to 750A (#1). At 3'35, there is already about 500 Angstroms deposited and the deposition rate is around 45A/s. The window is already completely aluminized to the edges, preventing you to see inside.
  • When thickness reaches 580 Angstroms, depress the "decrease" button until evaporation rate is 0 and current is 0. A thickness of 650 Ang. on the coating monitor will correspond to a maximum thickness of 1000 Ang. in the middle area of the mirror.
  • At 4'00, the game is over, the thickness monitor should be near 800A. Press "Off". TOTAL TIME of power in filaments: 4 min, TOTAL TIME aluminizing: 45sec.

B.35. Let the filaments cool for 15 min.

B.36. TURN OFF all gauges control.

B.37. CLOSE high vacuum valves 1 and 2. CLOSE throttle valves 1 and 2. TURN OFF liquid level controllers.

B.38. CLOSE foreline valves 1 and 2. TURN OFF roughing pump. TURN ON holding pump.

B.39. When pressure in holding lines is below 20 microns, OPEN Holding valves 1 and 2. TURN OFF Diffusion pumps.

B.40. Release slowly for 1min air to the chamber (big manual valve in the back) then open it completely. Let air get in 20min.

B.41. UNPLUG the LN2 cylinders and clean the lines with compressed air.

B.42. When the chamber is at atmospheric pressure, disconnect the head of the roughing pump and open the chamber. Remove test plates and mirror. Close the chamber.

B.43. When the diffusion pumps are cool to touch, CLOSE holding valves 1 and 2. TURN OFF Holding pump. TURN OFF control power. TURN OFF key switch.

B.44. TURN OFF air and water to diffusion pumps. CLOSE big manual air release valve.

 

** The goal is to deposit a layer of Aluminium of 950 Angstroms (±5%) of thickness on the glass (the chamber is capable of ±10% thickness uniformity when using the outer array only). Above 1000 Ang. thick, the coating will start showing more and more surface roughgness which will increase scatter. A thickness of at least 500 Ang. is required to maintain the transmission through the coating below 0.1%. The faster the evaporation rate is better because the vapour of Aluminium doesn't get much time to be contaminated with oxygen or other residual gases in the chamber. A rapidly deposited coating is more compact and show less surface roughness. Once the Al coating enters in contact with air, a 30 to 40 Angstroms-thick layer of Aluminium oxide (Al2O3) will quickly form on the surface and protect the Aluminium from tarnishing. This layer also hardens tremendously the Al (Knoop hardness of 2100 kg.mm-2 compared to 140 kg.mm-2 for bulk Al). Final experiments to check quality of aluminizing: look for water marks; put some sticky tape on edge of mirror and watch for aluminium peeling off; look at bright incandescent bulb from the edge of the mirror through the glass: the bulb should barely be visible. Inside the chamber, check the status of all the filaments : broken ones, clips not melted, eventual short circuits,... which will give you a better idea of the thickness uniformity.

 

Note: chemical processes in the obtention of the vacuum

  • Low vacuum: the gas is like the atmosphere and the number of molecules of gas is large compared to what is covering the surfaces. The process rarefy the existing gas from atmospheric pressure to 10-2 torrs.
  • High Vacuum: what is left is 70 to 90% of water vapour. Most of gas molecules are located principally on surfaces and their mean free path equals or is greater than the dimension of the enclosure. Vacuum goes from 10-3 to 10-7 torrs.
  • Glow discharge won't work at high vacuum because there is no conductive medium (gas). Inlet of ambient air (rather than dry air) is good because oxygen is reactive and helps removing the hydrocarbons.

Schematic diagram of the diffusion pump

pump0.jpg [15]

Figure 11


Maxime Boccas, last revision on 2th of August 2000.

Aluminizing images

alum50.jpg [16] The mirror is being washed before removing the coating.
alum20.jpg [17] The mirror is ready for a good wash after staying one year without any cleaning!
alum40.jpg [18] The mirror is being washed before removing the coating.
alum10.jpg [4] The old aluminium coating has been removed and shows the bare glass (Cervit).
alum30.jpg [19] The aluminizing chamber is opened, the new coating inspected and the mirror is going to be moved back to its cell.
alum60.jpg [2] The TCDT (Tololo Coating Dream Team) celebrating with champagne (no pisco). Standing up from left to right: Oscar Saa, Jorge Briones, Gale Brehmer, Mario Gonzalez, Eduardo Huanchicay (in the back) and Eduardo Aguirre (in front), Maxime Boccas, Gabriel Perez (at the bottom).

 

Materials

List of materials for the washing of the 4.0-m mirror

  • 1 kg orvus soap 0.5 kg cotton
  • 1 case Kimwipes XL lint free
  • 1 box Balzers optical cotton cloth
  • 1 cylinder dry-nitrogen
  • 100 liters Double distilled water
  • 6 liters HCl
  • 1/2 kg CuSO4
  • 2 kg KOH
  • 10 dust masks
  • 6 goggles
  • 6 respirators
  • 6 pairs heavy rubber gloves
  • 50 pairs talc free surgical gloves
  • 6 pairs rubber boots
  • 6 rain plastic suits
  • 1 box pH test paper
  • 1 inline tap water filter feeding 3 hoses
  • 6 clean plastic buckets for soap and water

Thickness monitor

SIGMA SQM160 thickness monitor operation

  • Connect sensor to feed-through bolt in chamber wall with a microdot in-vacuum cable).
  • Connect the oscillator input (labelled 'feedthrough') with the small BNC cable to the feed-through bolt connector.
  • Connect the oscillator output (labelled 'instrument') to Channel 1 plug in the rear panel of the monitor.
  • Turn ON the monitor. A permanent green light shines in the 'crystal status #1' (if the led blinks, it means that there is a poor connection: check the cables and that the sensor holder is tight on the support).
  • Breath gently on the sensor: the thickness should go up and down as water vapour deposits on it and then evaporates. That is the test to check that it works (do it before you close the chamber!).
  • Press 'Program' to check the coating parameters (see p2-3 of the booklet): most important is Al density 2.73, Z-factor 1.08, Tooling 100 (or what you determined). Press 'Next' to scroll down the list.
  • Press 'Zero' to reset the counter when you start aluminizing in order to see easily have the elapsed time of the process.

 

 

Max Boccas, last revision on February 1st, 2001

4.0-m In-situ washing procedure

This operation is conducted every 6 months. This is a 5 people job: 2 persons are needed for the washing and at least 3 assistants (one near the mirror to pass on the goods and change the hose connections, one near the buckets to change the water pump from one to the other, and one on the M floor to open water and check water exhaust). Here follows a list of materials used, a description of the process and some images:

Materials:

  •  2-4 natural sponges rinsed thoroughly in washing machine (they need to be absolutely free of particles, have a close look)
  • 1 large yellow plastic bucket with 50l of bidistilled water. Put a lid on the bucket until the moment you use it in order to avoid contamination of the water.
  • 1 large yellow plastic bucket with 50l of tap filtered water for pump rinsing
  • 1 large yellow plastic bucket with 50l of WARM tap filtered water with Orvus soap (make a rather concentrated solution)
  • 2 NEW nitrogen gas cylinders with full pressure with 2 spraying guns each. The cylinders must be FULLY charged (high pressure) otherwise the drying is too slow and will leave some thin water streaks.
  • LIGHTS: it is fundamental to have an intense illumination all over the mirror to see very well what one is doing and detect all the residal spots while washing and especially drying. Install the 3 sets of double lamps used for the 4m Aluminizing (borrow them from Gale) and hopefully an additionnal one, in the 4 corners of the telescope guirder.
  • the balcony bracket and its 2 wooden boards (remove the interior M1 ring baffle before putting the balcony)
  • 2 white plastic suits
  • 1 submersible pump
  • 1 green water hose connected to the water exhaust down to a large bucket on the M floor
  • 1 yellow water hose connected to the filtered water tap on M floor and with a quick connect fitting at the other extremity
  • 1 yellow water hose connected to the pump and with a quick connect fitting on the other extremity
  • 1 yellow water hose with a T fitting and 2 secondary hoses equipped with water spraying guns
  • 4 nitrogen gas guns with corresponding hoses
  • 2 nitrogen gas cylinders
  • 2 pairs of surgical gloves
  • 'water' vacuum cleaner in case of problem in the exhaust pipe
  • a few boxes of Kimwipes drying papers

 

Process:

  • Count 1 hour for the preparation of all the materials and 2 hours for the washing process to be completely done.
  • DO NOT FORGET to seal very carefully with wide strong adhesive tape (Duct tape) the gap between the interior seal and the flange on the chimney above it (there are holes there and unless they are sealed definitively with silicone, they need special sealing to avoid water getting into the chimney and the mirror cell)
  • Beware that the grounding plate at the top of the mirror prevents locally the pneumatic seal to do good contact with the mirror. Do not spray water into that tiny area or put some tape to seal it.
  • Inflate the exterior and interior pneumatic seals. Inspect visually that they are making good contact with the mirror. Tape the interior seal connector and the temperature sensor to the chimney.
  • Remove the ring baffle around the chimney and install the balcony.
  • Clean separately any greasy spots or any other weird marks like bird drops that could contaminate later the sponges and the rest of the mirror.
  • Hose water down into the exhaust hole until water flows at the output of the pipe. Then start spraying tap filtered water on the mirror for a few minutes to remove the dust.
  • At any time, the person on the M floor should check that there are no water filtration through the chimney, rotator and mirror cell (you would eventually see water leaking to the ground).
  • Spray the Orvus solution and start sponging the mirror. One has to climb on the chimney to reach easily the top of the mirror
  • Rinse a few minutes with tap water and repeat the Orvus sponging. Rinse again with tap water.
  • Put the pump into the filtered water bucket and keep rinsing the mirror until no more Orvus solution comes off the hose. This is to clean off Orvus from the pump and the hoses.
  • Reconnect to the tap filtered water for a few minutes (there is much more pressure than the filtered water in the bucket so you can eliminate all traces of Orvus).
  • Put the pump in the bidistilled water and spray generously all over the mirror until you have about 10l left.
  • Start drying the mirror: one person on the chimney dries out the surroundings of the mirror with Kimwipes, then blows nitrogen gas from the top to the bottom until it reaches a sufficient height to keep drying from the balcony. Remove the sealing tape around the chimney and dry with paper the water in the holes. MAKE SURE you don't blow that water back onto the mirror as it will leave water marks (this 'trapped' water is not perfectly clean).
  • Dry as fast as possible always from top to bottom (high pressure is needed in the cylinders). One person on each side with one gun in each hand.
  • A third person needs to maintain wet the lower part of the mirror by keeping spraying bidistilled water, especially below the chimney, until the drying persons are ready to get to that area.
  •  TAKE A LOT OF CARE at the end of the process of removing with paper all water on the surrounding of the mirror to avoid late contamination. Use Kimwipes to suck the pool of water left at the botttom.

 

wash1.jpg [20] For our first in-situ washing test (august 00), we had not yet received the exterior inflatable seal so we used duct tape and plastic sheet.
wash2.jpg [21] Detail of how the balcony is attached to the chimney with a special bracket.
wash3.jpg [22] Detail of the interior inflatable seal (between mirror center hole and chimney) and its valve.
wash4.jpg [23] Water is being sprayed on the mirror.
wash5.jpg [24] A small pool is forming at the bottom and being sucked up immediately by a water pump.
wash6.jpg [25] Detail of spraying water.
wash7.jpg [26] Detail of drying the mirror with dry nitrogen.

 

Last revised on February 2, 2001, Maxime Boccas

Aluminizing and washing calendar

Al & W from (1975-2001)

 

W stands for washing; Al stands for aluminizing

 

Date 4.0-m 1.5-m 1.0-m 0.9-m Schmidt instruments
             
3-Oct-01   W M1        
4-Aug-01 W M1          
30-Apr-01   W M1        
9-Apr-01       W M1    
8-Mar-01 W M3 (TT box)          
1-Feb-01 W M1          
30-Jan-01 Al M2(F8)          
3-Nov-00   W M1        
2-Nov-00       W M1    
9-Oct-00 W M2(F8)          
6-Sep-00   W M1        
Aug-00 Al M1          
12-Apr-00       W M1    
21-Mar-00   W M1        
9-Nov-99           Al mirror(?) Hydra 4m
24-Aug-99   Al M2s        
31-May-99       Al M1+M2    
1-Dec-98     Al M1+M2      
25-Sep-98           Al col Hydra 4m
31-Aug-98   Al M1        
5-Aug-98 Al M1          
9-Dec-97       Al M2    
16-Jun-97       Al M1    
9-Apr-97   Al M1        
17-Aug-96  Al M2(F14)          
28-Jun-96 Al M1          
12-Jul-95   Al M2s        
25-Apr-95      Al M1      
18-Oct-94       Al M1    
10-Aug-94  Al M1          
23-Jun-94   Al M1        
29-Dec-93  Al M2(F8)          
14-Oct-92        Al M1    
5-Oct-92      Al M1      
8-Sep-92    Al M2s        
17-Jul-92  Al M1          
23-Oct-91    Al M1        
9-Sep-91           Al Cam Echelle 4m
30-Aug-91  A M2(F7.5)          
27-May-91 Al M1          
5-Aug-90       Al M1+M2    
13-Aug-90     Al M1+M2      
19-Jul-90          Al M1  
15-Nov-89   Al M1        
19-Dec-88   Al M2(F7.5)        
26-Aug-88           Al col spectro 4m
30-Aug-88  Al M1          
30-Jun-88        Al M1    
31-May-88     Al M1      
Jan-88          Al Newton  
18-Aug-86   Al M1        
21-Mar-86 Al M1          
28-May-85        Al M1    
7-May-85      Al M1      
12-Jun-84   Al M1+M2(F7.5)        
23-Aug-83      Al M1      
21-Jul-83 Al M1          
24-May-83   Al M1+M2(F13.5)        
24-Nov-82       Al M1    
3-Aug-82 Al M2(F8)          
13-Jul-82       Al M2    
8-Jun-82     Al M2      
13-Apr-82   Al M1        
8-Sep-81     Al M1      
15-Jul-81  Al M1          
14-Apr-81       Al M1    
25-Nov-80   Al M1        
20-Nov-80           Al 4m rotator mirror
28-May-80    W M2(F13.5)        
31-Aug-79  Al M1          
21-Aug-79     Al M2      
9-Jul-79   Al M1        
14-Jun-79          Al M1  
9-Apr-79   Al M2s        
13-Mar-79     Al M1      
21-Aug-78      Al M2      
14-Aug-78       Al M1+M2    
18-Jul-78   Al M1        
12-Jun-78 Al M1          
29-Nov-77          W corrector  
22-Nov-77       Al M1    
26-Oct-77     Al M1      
7-Jul-77           Al 4m rotator mirrors
29-Jun-77   Al M1        
30-Mar-77           Al Col. Spectro 4m
28-Sep-76       Al M1    
Jul-76   Al M2s        
Jun-76       Al M2    
18-Apr-76     Al M1      
13-Nov-75         Al M1  
Aug-75       Al M1    
Date  4.0-m  1.5-m 1.0-m 0.9-m Schmidt instruments

 

Small Telescopes Aluminizing

Small Telescopes Aluminizing Procedures

 

(chamber in the 1.5m dome)

This aluminizing chamber accepts mirror up to 1.5m in diameter and is used for all CTIO mirrors but the 4m M1.

Mirror washing: see 4.0-m procedure [27].

Picture of the 0.9m tel M1 mirror [28] after sitting 2 years in the tube without any cleaning and a picture of the washing [29] (aluminium removal with acid).

Note: units used are : 1.3x10-6 atm = 1 micron (Hg) = 1 millitorr (=0.13 Pa)

Vacuum sensors and controllers: on the main panel, there is an analog controller NRC725 (Norton) for the Varian 524-2 cold cathode (1 input, reading from 10-3 to 10-7 torrs), and for Varian 531 thermocouples (2 inputs: not used not working anymore!). On top of the panel, there is an extra analog controller NRC721 fot 531 thermocouples (2 inputs: one for the diffusion pump and one for the tank, both reading from atm to 5 microns). There is also a thermocouple in the roughing line, which can be used to check good funcionning of the pump (turn off the meter, unplug the diffuser sensor and plug into the roughing line). Calibration of that dual thermocouple meter must be checked: when rough vacuum is reached in the roughing line (close diffuser and tank valves and plug the sensor in the roughing line), press the 'fil MA' button and turn the 'fil adj' knob until the needle reads 200 on the top red scale.

We indicate the elapsed time for the aluminizing (rather slow in this example, the entire process could last as little as 3h00).

 

A. PREPARATION OF CHAMBER (90 min)

A.1. Put a roll of paper on the bottom of the chamber to step inside without leaving dirt.

A.2. Replace all the damaged filaments (there are 24 of them on a single diameter): they must be straight when loose (i.e. unstressed), not covered with aluminium (drops). See picture [30] of this operation inside the chamber. Press gently the filaments with the fingers : if they are stressed, they break immediately. Filaments can usually be used for 2-3 aluminizings. Handling of the filaments must be done with gloves to avoid their contamination with grease.

A.3. Prepare 96 aluminium clips and wash them in acetone. Hang 4 clips per filament : one on each loop. Again use gloves to hung the clips.

A.4. Check thoroughly cleanliness of chamber, wash walls with acetone and all the surfaces touched by hands or tools, check the chamber Orings (clean and lubricate)

A.5. Use mouth mask from the moment the mirror is dried out until the chamber is closed. Wash with acetone the inside of the mirror holding ring and attach to the mirror. Bring the mirror into the aluminizing tank (picture [31]). Wash with acetone the outside of the mirror holding ring and any spots touched by tools.

A.6. Install the thickness monitor: it must face the center of the filament array and be at the same distance from the filaments as the mirror (otherwise you need to enter a 'Tooling' coefficient in the thickness monitor meter. Check that it is working (see instructions [11])

A.7. Blow off laterally dust with dry nitrogen. This is a critical step to ensure that the mirror is absolutely free of dust. Close tank and seal it by tightening the 4 bolts. Add at least 2 C-clamps near the top (original bolts are missing).


B. ALUMINIZING (see picture [32] of control panel for location of all knobs)

B.1. 0 min: Connect black and orange power cables (filaments and glow discharge) on the side of the main control panel. Connect 2nd black cable from glow discharge controler to valve at the rear of the tank (on the air inlet tube). Connect main thick black power cable to wall plug.

B.2. Take the extremity of the black flexible exhaust water pipe outside the building.

B.3. Open the water circuit valve (green pipe), the valve to the roughing pump and to the thickness monitor.

B.4. Close the copper tube valve (water to diffuser) on the right side of the control panel

B.5. See all following elements described on picture 1 [33]. Connect the tank vacuum gauge. Close both valves (diffuser and tank): the handle of the tank (top valve) should be on the left side and the handle of the diffuser (bottom valve) should be on the right side.

B.6. Picture 2 [34]. Turn on the air compressor: it must show 100 lbs to be able to open/close the diffuser valve. Make sure the voltage wheels for glow discharge and filaments are at 0 and their respective controllers off.

B.7 Turn on the roughing pump, plug a thermocouple into the line and check the meter calibration (as described above) once good vacuum is reached (5 um). Reconnect the sensor to the diffuser and open the diffuser valve.

B.8. When diffuser gets to 5 microns, close diffuser valve and open tank valve. Some smoke is generated by the compressor for a few minutes: close the doors and turn on the room air extractor.

B.9. 25 min: When tank vacuum reads 200 microns, turn on the power knob on the control panel to start the glow discharge. Turn control knob to raise current progressively up to 500mA (about 700V) (picture [35]) and check the ionization color in the tank (watch out for arcs). The vacuum should drop around 30 microns. Maintain glow discharge for 20min. Decrease slowly to zero the voltage, switch off the controller and turn off the glow discharge power knob on the main control panel.

B.10. 1h35min: When tank is at 5 microns, close the 2 valves (diffuser and tank) and open the poppet valve (diffuser-to-tank valve) by switching on the appropriate knob (this makes a loud noise, don't panic!). Turn on the diffuser, the refrigeration and open the copper tube water inlet. Turn off the vacuum gauge and turn on the cold cathode on the first scale (10-3 to 10-5 torr).

B.11. Open the diffuser valve (so that the roughing pump starts pumping the tank through the diffuser).

B.12. 3h50min: the tank reaches 4x10-5. Check the thickness monitor [11] is ready to use. Turn on the filaments power on the main panel (left 'low voltage' knob) and on the filament controller.

B.13. Always reading the values of the current on the wheel -not on the small meter- (picture [36]), raise slowly to 35A and check for 2-3 min until the filaments color is uniform.

B.14. Raise to 45A for 1 min, then to 55A for 2 min: you should see the aluminium melt on the filaments (the filaments darken, then redden again). The aluminizing process is initiated and you should see some activity on the thickness monitor display.

B.15. Raise to 65A for 30sec: all clips should be melted and the filaments of a uniform color with a deposition rate around 15A/sec. Raise to 80A for 30 sec then 90A until thickness reaches 890A (maximum rate should be around 17A/sec). Decrease the power down to 0 in 10 sec. Switch off the power on the controller and on the main panel. Real aluminizing time should be around 1min30sec and final thickness should be around 950A.

B.16. Close the poppet valve, and turn off the compressor. Switch off the diffusion pump. Turn off the cold cathode and disconnect it.

B.17. Open a little the air inlet valve to the tank (with a piece of cloth on the pipe entrance). When diffuser cooling water exhaust is cool again, close the diffuser valve, turn off the roughing pump, the refrigeration and close the main water valve on the wall (leave the copper tube opened).

B.18. It will take about 20 minutes for the tank to be at atmospheric pressure so that you can open it (listen when air leaking in stops).

B.19. Inspect carefully the coating and look for water marks with intense light shining on the coating. Check visually that there is no transmnission through the coating (ie. thickness is adequate) nor tiny 'dust' holes in it by shining a bright light from behind the mirror. Do the adhesive tape test on the edge of the mirror to check the quality(adhesion) of the coating. Measure the reflected and scattered light in 3 different places to compare with data before aluminizing.

Total: 4h30min.

 

Max Boccas, last revision on February 1st, 2001

Small Telescopes In-situ washing procedure

Always check that you have plenty of pressure and reserve in the nitrogern gas cylinders before you start the washing process in order to make the drying fast and optimum.
 

1.5-m Telescope

Washing done in-situ. Remove chimney. Put telescope at ZD60deg to the south. Raise the platform as much as possible. Remove northern petal covering M1, install the pneumatic seal around M1 so that it protrudes about 20mm above the mirror edge, and inflate it. Seal VERY carefully the space between the chimney base and the inner hole of the mirror with duct tape. If you don't remove the instrument from the telescope, make very sure all the seal are effective! Prepare a warm and highly concentrated solution of orvus soap and water (lot of foam). Rinse with tap water and hose. Use the vacuum cleaner to suck water. Contact-wash with the natural sponge and the soapy solution, and try to maintain the glass wet and covered with foam for 5 minutes (the idea is to unstick the dust and also degrease). Rinse with tap water. As usual, observe how the water is flowing on the surface, any abnormal surface tension showing up will indicate a residual grease that has to be removed. Never let the mirror dry. Finally rinse with bidistilled water (about 6-8 liters needed in small 1 liter bottle) and dry with at least one high pressure nitrogen gas gun (check you have plenty of pressure in the nitrogen cylinder before you start). Dry carefully with KImwipes all the water drops remaining around the mirror. With lot of care to avoid trapped water jumping, remove the inflatable seal and unstick the adhesive tape.

 

0.9-m Telescope

Remove the cell from the tube, incline it by 5-10deg. Mirror stays in cell. Seal the inner hole by forming around it a 2"-high cylindrical wall with duct tape and seal the outer diameter with the special plastic round skirt and duct tape. Contact-wash with sponge and soapy solution. Suck the water from the lower side of the mirror with vacuum cleaner. Dry with 2 nitrogen gas guns. The entire process takes about 3-4 hours (the washing itself takes 30min at most) and can be done without engineering time as the telescope collimation is not affected once the cell is reinstalled.
 

 

4m CFADC top surface (sol-gel AR coated):

Bring the telescope at D=30deg North (same position as M1 cleaning) and work inside the chimney with proper working light. First blow off dust with dry nitrogen. Only about 100ml of dichlorodimethylsilane (DDMS) is used in a goose-neck plastic bottle. A large plastic bag is taped on the lower half of the cell perimeter to force water to flow directly into it. Two Orings seal the top element with the cell and the cell itself so that no liquid can get inside the corrector assembly (see CH2903-E003) but you still need to take some care. Use some Kim-wipes towels placed at the bottom of the lens to suck liquid as it flows down the glass, and change the towels once they are too wet. Dry off with nitrogen gas. The entire process takes at most 1h.

IMPORTANT: solgel coatings are hygroscopic (they absorb water) and deteriorate over time. Waterproofing is achieved by rinsing the coating with a solution of DDMS with ethanol to a concentration of 5 parts per million. Alcohol is not good because it will wash off the DDMS and leave the solgel unprotected against humidity! Be careful, DDMS is very nasty stuff. Prepare the dilution in 2 steps: first 0.1ml of DDMS into 100ml of ethanol and then 0.5ml of this solution into 100ml of ethanol.
 

 

M3 in F/14 TT BOX (gold coated):

Remove cell from box, remove mirror from cell (it is pushed against a reference corner with 2 plungers), and wash in a soapy and warm bath, rinse and dry with nitrogen gas. Some minor tilt adjustment is needed when its cell returns into the box (do it with Osiris pupil imagery).

 

 

Reports

0.9-m Washing 5Feb09 [37]

1.0m Washing 5Feb09 [38]

 

April 11, 2001, Maxime Boccas

 

Products for aluminizing and cleaning

  • R.D. Mathis Co. [39] (filaments for the 4m chamber)
  • R.H. Cheney Inc [40]. (sells 99.99% Al clip evaporant, at only a few dollars for a bag of 10,000!, tel 508 226-7300)
  • Sigma instruments [41] (thickness monitors)
  • Midwest Tugsten Service [42] (about metal evaporation)
  • Osram Sylvania emissive products [43] (filaments for the 1.5m chamber, tipe 116586, code X238D)
  • Newport Thin Films Laboratory [44] (optical coatings)
  • Varian vacuum technologies [45] (vacuum gauges and others)
  • Thin Film Technologies Inc. [46]
  • Denton Vacuum [47]
  • United vacuum materials [48] (evaporation and sputtering)
  • Sotware Spectra Inc [49] (thin film software)
  • AVS, the Science and Technology Society [50] (buyers guide)
  • Target materials Inc. [51]
  • Chemical elements on the web: tungstene [52]

 

  • Vatran C02 cleaning [53]
  • www.co2clean.com [54]
  • DMO reflecto-scatterometers [55]
  • Schmitt Industries [56] (Micro-Scan reflecto-scatterometer)
  • Minolta's spectro-photometers [57]

 

  • Photonics directory [58]: metallic coatings buyers guide
  • Sputtering targets suppliers [59]
  • Atkinson Thin Film systems [60]
  • Barr Associates Inc. (filters) [61]
  • NASA materials databases [62]
     

Source URL (modified on 04/05/2011 - 14:44): http://www.ctio.noao.edu/noao/content/Aluminizarion-Washing

Links
[1] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/blanco-al-tests-2009-summary.xls
[2] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/alum60.jpg
[3] http://www.ctio.noao.edu/noao/content/materials
[4] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/alum10.jpg
[5] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/inchamb2.jpg
[6] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/inchamb.jpg
[7] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/glow.jpg
[8] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/filament.jpg
[9] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/4mpanelvac.jpg
[10] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/panel0.jpg
[11] http://www.ctio.noao.edu/noao/content/thickness-monitor
[12] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/vacback.jpg
[13] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/vacback2.jpg
[14] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/glowdis.jpg
[15] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/pump0_0.jpg
[16] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/alum50.jpg
[17] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/alum20.jpg
[18] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/alum40.jpg
[19] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/alum30.jpg
[20] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash1.jpg
[21] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash2.jpg
[22] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash3.jpg
[23] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash4.jpg
[24] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash5.jpg
[25] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash6.jpg
[26] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/wash7.jpg
[27] http://www.ctio.noao.edu/noao/content/40-m-aluminizing-procedure
[28] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/dirty.jpg
[29] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/cleaning.jpg
[30] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/incam.jpg
[31] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/mirincam.jpg
[32] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/tabmain.jpg
[33] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/chambert1.jpg
[34] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/chambert.jpg
[35] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/tabglow.jpg
[36] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/tabfila.jpg
[37] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/0.9m_washing_5Feb09.pdf
[38] http://www.ctio.noao.edu/noao/sites/default/files/telescopes/1.0m_washing_Feb09.pdf
[39] http://www.rdmathis.com/index.html
[40] http://www.rhcheney.com/index.html
[41] http://www.sig-inst.com/
[42] http://www.tungsten.com/mtsprod.html
[43] http://www.sylvania.com/
[44] http://www.newportlab.com/products.htm
[45] http://www.varianinc.com/cgi-bin/nav?/products/vacuum/
[46] http://www.thinfilmtech.net/
[47] http://www.dentonvacuum.com/
[48] http://www.uvmab.com/
[49] http://www.sspectra.com/index.html
[50] http://www.aip.org/avsguide/index.html
[51] http://www.targetmaterials.com/
[52] http://www.webelements.com/tungsten/
[53] http://www.vatran.com/index.html
[54] http://www.co2clean.com/
[55] http://www.dmo.be/
[56] http://www.schmitt-ind.com/
[57] http://www.konicaminolta.com/instruments/index.html
[58] http://www.photonicsonline.com/
[59] http://metals.about.com/od/sputteringtargets/Sputtering_Target_Suppliers.htm
[60] http://www.sputtercoat.com
[61] http://materion.com/Businesses/Barr%20Precision%20Optics%20and%20Thin%20Film%20Coatings.aspx
[62] http://setas-www.larc.nasa.gov/LDEF/MATERIALS/mat_databases.html