Date:
Wednesday, March 8, 2023
Telescope:
CTIO 1.5-m
Submitter:
Jacqueline Seron
Submitter Email:
Observer Support:
None
Night Assistant:
Jackie Seron
Problems:
Mac pc for data acquisition is slow, it needs a reboot. I’ll do it tomorrow.
Telescope would drift to the West at times when not guiding (didn’t take notes on the positions).
I run calib2 but the first ThAr image looks like a zero. It seems it was stuck trying to move the slicer. I re-started CHIRON GUI and run calb2 again. It seemed to work this time.
Program:
Target of Oportunity:
No
Time Observed:
10.0
Time Engineering:
0.0
Time lost technical:
0.0
Time lost weather:
0.0
Time lost other:
0.0
Primary Investigator:
Todd Henry
Email:
Institution:
SMARTS
Instrument:
CTIO 1.5m Echelle
Organization:
SMARTS
Total Program Time:
10
Total Times | Time Observed | Time Engineering | Time lost technical | Time lost weather | Time lost other | Total Program Time |
---|---|---|---|---|---|---|
2023-03-08 | 10.0 | 0.0 | 0.0 | 0.0 | 0.0 | 10 |
Back to the 1.5m telescope after more than 8 years! (my last night was October 25, 2014). Todd joined me at the beginning and helped me out. I'm a little rusty, but I think I caught on with the system.
Clear night and program was almost completed, one target skipped: HD152268.
I re-took some ThAr lamps because they looked like zeroes at the beginning of the night. Maybe I did something wrong, I’m sorry.
GAI1105+0057_gap faint star for guider and there was no other star available.
No data on dewar pressure. The detector is off.
A note: In the queue ThAr lamps it says 4 sec exposure but when are actually 1 sec exposures.