Spectroscopic Mode Overview
OSIRIS has three spectroscopic modes: low-resolution (R~1200) and
high-resolution
(R~3000) long-slit modes plus a low-resolution cross-dispersed mode
(R~1200)
with a short slit. Separate slits are provided for each mode as
described
below.
OSIRIS is put into spectroscopic mode by retracting the
prefilter
wheel and pupil mask, and then flipping the grating into the beam in
place
of the imaging flat. This operation requires approximately 2 minutes to
complete. The imaging flat may be flipped back into the beam rapidly (3
seconds) to provide a quick-look Acquisition
Mode to center the object on the position of the slit (either by
taking
an image of the field or viewing through the slit).
This section describes:
Spectroscopic observing strategies and techniques (including target
acquisition
and calibration issues) are discussed in the section on Spectroscopic Observing.
OSIRIS Spectroscopic Modes
OSIRIS has three spectroscopic modes:
-
Low-Res
-
R~1200 single-order long-slit spectroscopy covering all of the J, H, or
K bands in a single
spectrum. The Low-Res long slit covers the entire width
of the field of view of the f/2.8 camera at a fixed width. The grating
tilt should be set at 950 in low-res
mode.
-
High-Res
-
R~3000 single-order long-slit spectroscopy covering most of the J, H, or
K bands. The wavelength
region is adjusted by tilting
the grating. The High-Res long slit covers approximately 80 arcseconds
by 0.48 arcseconds.
-
X-Disp
-
Low-Resolution (R~1200) multi-order cross-dispersed
spectroscopy with a short slit and the f/2.8 camera that covers all of
J, H, and K bands simultaneously in adjacent
orders. The cross-dispersed
slit is only ~15% the height of the low-res long slit at a fixed width.
All modes use a single diffraction grating blazed at 6.6-microns in
first
order, which places the K-band
in 3rd order, H-band in 4th order, and J-band
in 5th order. A grism and JHK
filter (integrated into the short slit) are
used to cross-disperse orders 3,4, and 5 (and part of order 6) across
the
detector.
The High- and Low-resolution modes are selected by changing
the
camera and slit, and using one of the filters as an order sorter.
Cross-dispersed
("X-Disp") mode is selected by introducing the cross-dispering grism
(in
the filter wheel) and short slit into the beam. This slit has an
integrated
JHK filter to provide blocking outside of the bands of interest. The
table
below summarizes the camera, slit, and filter configuration used by
each
of the OSIRIS spectroscopic modes.
Table 1: OSIRIS Spectroscopic
Configurations
Mode |
Camera |
Slit |
Filter |
Low-Res |
f/2.8 |
0.34mm x 63mm
(LoRes Long) |
J, H, K
or LPK |
High-Res |
f/7 |
0.14mm x 24mm
(HiRes Long) |
J, H, K
or LPK |
X-Disp |
f/2.8 |
0.34mm x 8.6mm
(XDisp Slit) |
Grisma |
The order separation filters are as follows
-
J Filter for 5th order (center ~1.25-microns).
-
H Filter for 4th order (center ~1.65-microns).
-
K Filter for 3rd order (center ~2.2-microns).
-
Using the Long-Pass K (LPK) Filter in 3rd order extends the red
coverage
a little past 2.4-microns (until the thermal background and atmosphere
clobber you).
The following table summarizes the basic properties of the OSIRIS
spectroscopic
modes. The properties of the slits depend on the telescope, and are
summarized
in below.
Table 2: Spectroscopic Parameters
Mode |
Filter
(order) |
Center
Wavelength (Å) |
Linear Dispersion
(Å/pix) |
Wavelength Range
(Å) |
Resolution
(2 pix) |
Low-Resa |
J (5) |
12500 |
5.525 |
2500 |
1200 |
|
H (4) |
16500 |
6.906 |
3000 |
1200 |
|
K (3) |
22000 |
9.208 |
3800 |
1200 |
High-Res |
J (5) |
selectb |
2.208 |
1435 |
3000 |
|
H (4) |
selectb |
2.762 |
1795 |
3000 |
|
K (3) |
selectb |
3.683 |
2394 |
3000 |
X-Disp |
JHKc
(5-3) |
n/a |
(low-res)d |
10000-24000
(3.5 orders) |
(low-res)d |
Table Notes:
-
a) Low-Res mode central wavelengths are given for the
optimal grating
tilts (see Table 4) that center the entire band
within
the limits of the filter transmissions.
-
b) Center wavelength is selected by tilting the grating (see
below),
within limits prescribed by the filter transmissions and the physical
limits
of the grating travel.
- c) The "JHK"
filter used in X-Disp mode is integrated into the cross-dispersed
slit, and not in the filter wheel proper (which is occupied by the
cross-dispersing
grism).
-
d) Dispersion and resolution depend on the order, and are
the same
as for the individual low-resolution orders given in the table.
-
e) Because of vignetting by the f/7 camera, the true
coverage for
extracted spectra of a given tilt should be restricted to pixels
between
about 250 and 900 even though there will be dispersed light outside
this
range.
[ Top of Page | Contents
]
OSIRIS Slits
Slit widths and heights are fixed (pre-cut masks) as described in Table
3 below, and depend on which telescope you are using.
Table 3: Slit Properties
|
Physical |
CTIO 4-m |
SOAR |
Slit |
Width |
Height |
Width |
Height |
Width |
Height |
Low-Res Long Slit |
0.34mm |
63.0mm |
1.2" |
216" |
1.0'' |
191" |
X-Disp Slit |
0.34mm |
8.64mm |
1.2" |
30" |
1" |
27" |
High-Res Long Slit |
0.14mm |
24.0mm |
0.5" |
81" |
0.42" |
72" |
The Low-Res and High-Res long slit heights span the entire
unvignetted
imaging fields of the f/2.8 and f/7 cameras, respectively. The X-Disp
slit
is only 14% the length of the long slit so that the different orders
can
be stacked on the array without overlap.
Slit Orientation
By default, the instrument rotator = 0 degrees
when OSIRIS is mounted, and the slit is E-W
on the sky.
You can then check
the accuracy of the PA by doing a slitscan in Acquisition mode, measuring the
stellar
Y pixel position.
[ Top of Page | Contents
]
Selecting Wavelengths
The wavelength region is selected by tilting the diffraction grating.
How
much you need to tilt for a given configuration is described for each
mode
below.
Low-Res Long-Slit Mode
Since all of the given band is on the detector at once in
low-resolution
long-slit mode, the optimal grating tilts to center each band have been
measured in advance and are summarized in Table 4.
In general, it is not necessary to set the grating tilt to other than
these
settings.
The "center" here is defined as the wavelength at row
575
(the center of the f/2.8 imaging mask) in a vertical cut made along
column
515 (the center of the low-resolution long slit). The slit is curved as
seen on the detector such that the central wavelength of an object
shifts
to the RED as its position moves
toward
the end of the slit from the center. The shift is about 10 pixels from
center to end.
Table 4: Low-Res Mode Grating Tilts
(1999 Jan 25)
Filter |
Order |
Tilt |
Central
Wavelength |
Prospero
Command |
J |
5 |
950 |
12500Å |
|
H |
4 |
950 |
16500Å |
|
K |
3 |
950 |
22000Å |
|
If adjustments are required, the general rules of thumb for
tilting
the grating in low-res mode are as follows:
-
POSITIVE STEPS make the central wavelength BLUER,
and move lines toward LOWER pixel values (on the IRAF display)
by
approximately 1 PIXEL/STEP.
-
NEGATIVE STEPS make the central wavelength REDDER,
and move lines toward HIGHER pixel values (on the IRAF display)
by approximately 1 PIXEL/STEP.
Cross-Dispersed Mode
Since all of the principal orders (J through K) are on the detector at
once in X-Disp mode, the optimal grating tilt for this mode has been
measured
in advance. We do not recommend using a different grating tilt in
X-Disp
mode unless there have been serious mechanical changes in the
instrument
(at which point the support scientist should be immediately informed of
this).
-
X-Disp Grating Tilt: 950
-
Prospero Command: XDSPEC
In general, the optimal tilt for X-Disp mode is selected so as to best
center 4th order (H-band) on the array. The adjacent orders (J and K,
respectively),
sort themselves out across the detector with minimal losses off the
edges
of the field of view (most of these regions are in bad atmospheric
windows,
anyway).
See the Cross-Dispersed
Mode Order Map for details of how the orders are arranged across
the
array and descriptions of some of the other stuff you might see.
High-Res Long-Slit Mode
It is necessary to tilt the grating in order to cover the alternate
ends
of the JHK bands in high-resolution long-slit mode. The "center"
for high-res mode is defined approximately at pixel row 575 (the center
of the high-res long slit). Since the image of the slit is slightly
curved,
the object spectrum is shifted to the RED
(the wavelength at constant y pixel gets bluer) as you move
out from the center of the slit to the end by a little over 4 pixels
(8-12Å,
depending on the order).
The grating tilt can generally place the wavelength of interest
within
~1 pixel of the center (as defined above).
In general, the rules of thumb for adjusting the grating in
high-res
mode are as follows:
-
POSITIVE STEPS make the central wavelength BLUER,
and move lines toward LOWER pixel values (on the IRAF display)
by
approximately 2.5 PIXELS/STEP
-
NEGATIVE STEPS make the central wavelength REDDER,
and move lines toward HIGHER pixel values (on the IRAF display)
by approximately 2.5 PIXELS/STEP
-
For a tilt of 950, the J, H,
and K wavelengths of
pixel row
575 are 1.30181, 1.62726, and 2.16968
microns, respectively.
- For J, H, and K, the motion in wavelength is
approximately 5.520e-4, 6.905e-4, and
9.208e-4 microns/step, respectively.
[ Top of Page | Contents
]
Spectroscopic Performance
Sensitivity
Accurate sensitivity numbers have not yet been measured.
OSIRIS was designed to work at f/17, hence the
collimated
beam overfills the grating at f/14.5 resulting in a loss of throughput
(presently estimated up to factor of ~2). It is hoped that for many
programs,
the combination of increased wavelength coverage at higher spectral
resolution
(relative to the IRS) and increased visibility between the OH sky lines
for the XD mode will make up for this.
Grating Stability & Repeatability
Tests show that once a grating tilt has been selected it will be stable
to within 0.05 pixels RMS if not moved. If you are only going to be
using
one spectroscopic setting for a night (e.g., only taking
cross-dispersed
spectra), it is best to set the grating once and then leave it alone,
unless
you have evidence that the grating has been moved.
If you change the grating tilt, even setting it to the same
tilt
as it was before, this operation involves some mechanical imprecision
in
the repeatability of the encoders. Typically, if you ask the instrument
to reset the grating to its home position and then tilt it to a desired
setting, the grating will be driven to within ~1 pixel RMS in high-res
mode, and ~0.4 pixels RMS in the low-res and X-disp modes. The
non-repeatability
in tilt is due to the discreteness of the tilt encoders (recall that 1
encoder step moves a line by 2.5 pixels in high-res mode, and 1 pixel
in
low-res mode). You will probably have to take new standards and
wavelength
calibration frames if you move the grating and restore it to the same
position.
New flats will probably not be necessary.
There are occasional mechanical glitches when flipping between
spectroscopic and acquisition modes that can introduce approximately
0.3
pixels of shift in the high-res mode, and 0.1 pixels in the low-res
mode.
It is not repeatable, and appears as a slight jump in the spectral
lines
on close inspection. If you see a "large" jump of 2-3 pixels, you
should
reset the flip by typing ACQMODE then SPMODE to "cycle" the
grating/mirror
flip mechanism. These glitches are more likely to occur if the
instrument
has not yet reached equilibrium after being filled following warm up,
or
if it has run out of cryogen and is starting to warm up.
Flexure
To be added.
[ Top of Page | Contents
]
Updated: 2005 September 07 [rdb]