Next: RESTART AFTER HEURIKON DSP
Up: WILDFIRE RESTART PROCEDURES
Previous: TOTALLY STALLED SYSTEM RESTART:
If the power to the instrument was interrupted but the white Heurikon DSP box
in the computer room remained powered up and the computer was not rebooted:
- In the Instrument Control window, enter startwf following the
interactive startup procedure, turning the instrument power ON when requested
- 1.
- Press <return>
- 2.
- Answer n to the DSP box power question
- 3.
- Answer y to the instrument power question
- 4.
- Turn on the instrument power now: first DCU, then ACU.
- 5.
- Answer y to the instrument power being on
- 6.
- Answer y to wanting windows
- 7.
- Answer IRS (note it must be all caps) to the instrument selection
- Enter setup irs, answering 'y' to question about detector activation
- Enter the proper bias for the detector (nominal 0.55)
- Enter puse [parameter_file] to restore the parameter file active before the crash
- Verify ``pixel_dir'', ``header_dir'', and ``nextpic'' with plist or
ped to make sure data will still be going to the desired location and will not overwrite existing data
- Reload any scripts with the source command
- Continue observing
Next: RESTART AFTER HEURIKON DSP
Up: WILDFIRE RESTART PROCEDURES
Previous: TOTALLY STALLED SYSTEM RESTART:
robert blum x297
1998-04-25