If WILDFIRE has crashed (Instrument Status window has vanished and could not
be found by above procedures), and/or the ``[hostcomputer]'' prompt has returned
to the Instrument Control window, the following steps within the Instrument
Control window should restore operation:
- If the ``%'' prompt is present, first enter exit to cleanly exit
the processes and get the ``[hostcomputer]'' prompt
- At the ``[hostcomputer]'' prompt, enter: restart CIRIM, answering `y' to question about windows (note CAPS)
- Answer y to the question about motor control
- Enter fwinit to initialize the filter wheel
- Enter filter blank to select the dark filter
- Enter setup cirim, answering 'y' to question about detector activation
Enter the proper bias for the detector (nominal 0.90).
- Enter puse [parameter_file] to restore the parameter file active before the crash
- Verify ``pixel_dir'', ``header_dir'', and ``nextpic'' with plist or
ped to make sure data will still be going to the desired
location and will not overwrite existing data
- Reload any scripts with the source command
- Continue observing
[NOTE: If the power to the instrument and/or the Heurikon DSP box in the
computer room has been interrupted or the computer has been rebooted, this
procedure may not be sufficient. See below for more specific procedures]
robert blum x297