Closed for the night because of COSMOS camera stage motor problem.
COSMOS is in the Coude room on the instrument cart and open. During the afternoon we were able to move the camera stage when the instrument is horizontal. We verified the change by take slit images and measuring the FWHM of the slit. When we moved the instrument to a vertical position, we could hear the motor trying to move, but the camera would slip. We will continue with repairs tomorrow.
| Total Times | Time Observed | Time Engineering | Time lost technical | Time lost weather | Time lost other | Total Program Time |
|---|---|---|---|---|---|---|
| 2018-05-24 | 0.0 | 0.0 | 11.25 | 0.0 | 0.0 | 11.25 |