id |
sensor_type |
location |
description |
101 |
Surface Temperature RTD |
M1 coated surface North (Before 2018 shutdown: edge glued 5cm below surface) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor (Before 2018 Shutdown: Omega F2020-100-B-100) |
102 |
Surface Temperature RTD |
M1 coated surface South (Before 2018 Shutdown: on edge glued 5cm below surface) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor (Before '18 Shutdown:F2020-100-B-100 on 1x1"Al) |
103 |
Surface Temperature RTD |
M1 coated surface East (Before 2018 shutdown: at edge glued 5cm below surface) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor (Before '18 shutdown:F2020-100-B-100 on 1x1"Al) |
104 |
Surface Temperature RTD |
M1 coated surface West (Before 2018 Shutdown: on edge glued 5cm below surface) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor (F2020-100-B-100 on 1x1" Al block) |
105 |
Air Temperature RTD |
M1 Air South (Before 2018 Shutdown: it was glued to coating) |
RTD Omega PR-11-2-100-1/8-6-E (Before 2018 Shutdown: 1PT100FR828 thin film in ceramic body) |
106 |
Air Temperature RTD |
M1 Air North (Before 2018 Shutdown:M1 surface glued on north) |
RTD Omega PR-11-2-100-1/8-6-E (Before 2018 Shutdown: F2020-100-B-100) |
107 |
Bottom Temperature RTD |
M1 bottom substrate South (Before 2018 Shutdown: M1 surface glued on south) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor (Before 2018 Shutdown:F2020-100-B-100) |
108 |
Air Temperature RTD |
M1 Air East (Before 2018 Shutdown: glued on surface east) |
RTD Omega PR-11-2-100-1/8-6-E (Before 2018 Shutdown:F2020-100-B-100) |
109 |
Air Temperature RTD |
M1 Air West (Before 2018 Shutdown: surface glued on west) |
RTD Omega PR-11-2-100-1/8-6-E (Before 2018 Shutdown: Omega F2020-100-B-100) |
110 |
Bottom Temperature RTD |
M1 substrate bottom North (New since 2018 Shutdown) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
111 |
Bottom Temperature RTD |
M1 substrate bottom West (New since 2018 Shutdown) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
112 |
Bottom Temperature RTD |
M1 substrate bottom East (New since 2018 Shutdown) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
203 |
temperature |
Chimney ladder 1m above cell |
RTD PT100 Omega 1PT100FR828 in a 1x1x6 cm3 aluminium cylinder |
211 |
Surface RTD |
Cover (close to Petalos) |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
212 |
Surface RTD |
Mount structure East |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
213 |
Surface RTD |
Mount structure North |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
214 |
Surface RTD |
Mount structure West |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor |
301 |
temperature |
Top ring north side |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
302 |
temperature |
Top ring south side |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
303 |
temperature |
Top ring east side |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
304 |
temperature |
Top ring west side |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
305 |
temperature |
Decam Top Cap |
Omega SA2F-RTD-3-100-A-40 FLEXIBLE MOLDED RTD SENSOR |
401 |
temperature/rh/dewpoint |
Old console level close to west stairs |
Omega iTHP-2 industrial probe |
402 |
temperature/rh/dewpoint |
Clean Room |
Omega iTHP-2 industrial probe |
403 |
temperature/rh/dewpoint |
Coude Room wall |
Omega iTHP-2 industrial probe |
404 |
Air Temperature RTD |
East Wall side of Cass-Cage |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E on Channel 3 |
405 |
Air Temperature RTD |
West Wall side of Cass-Cage |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E on Channel 4 |
501 |
temperature |
Outside dome north at M level in a shelter |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
502 |
temperature |
Mirror Cooling: cooler input glycol pipe |
RTD PT100 |
503 |
temperature |
Mirror Cooling: air temperature just after cooler |
RTD PT100 |
504 |
temperature |
Mirror Cooling: air input in M1 cell at Dec Axis |
RTD PT100 Omega PR-10-2-100-1/4-6-E |
505 |
Surface Temperature RTD |
Mirror floor above glycol coil |
RTD Omega SA2C-RTD-3-100-A-40 Flexible Molded sensor(Bef 2018 Shutdown: 1PT100FR828 in Al 8x8x2cm3) |
506 |
voltage dc |
Mirror Cooling: v/p transducer feedback terminals |
Mamac EP-310; terminals 5(+) and 6(-) |
507 |
voltage dc |
Mirror Cooling: cooler input glycol pipe |
Omega Paddle Wheel Flow Sensor FP-5300 connected to Flowmeter FPM-800 terminals 3 and 4 |
508 |
Floor temperature |
C Floor (Old Console) above glycol coil |
RTD PT100 Omega 1PT100FR828 thin film in ceramic body mounted in Aluminium Block 8x8x2 cm3 |
509 |
temperature |
Mirror Cooling: cooler output glycol pipe |
Omega SA1-RTD-B surface RTD PT100 class B |
510 |
temperature |
Dome platform close to crane |
Omega zED wireless temperature End Device (zCDR coordinator) |
601 |
wind speed/wind dir |
Old console above 1m over roof |
Vaisala WS425 ultrasonic wind sensor rs-232 |
701 |
temperature |
Glycol pipe at Oil cooling system on P floor |
RTD PT100 |
702 |
temperature |
Oil supply NE radial hydrostatic pad |
RTD PT100 |
703 |
temperature |
Oil return NE radial hydrostatic pad |
RTD PT100 |
704 |
temperature |
Oil Cooling: Glycol supply pipe |
RTD PT100 |
705 |
temperature |
Oil Cooling: before cooler glycol pipe |
RTD PT100 |
706 |
temperature |
Oil Cooling: oil pipe before cooling |
RTD PT100 |
707 |
temperature |
Oil Cooling: oil pipe after cooling |
RTD PT100 |
708 |
voltage dc |
Oil Cooling: V/P Transducer feedback terminals |
n/a |
709 |
voltage dc |
Oil Cooling: glycol input line before cooler |
Omega Paddle Wheel Flow Sensor FP-5300 connected to Flowmeter FPM-800 (terminals 3 and 4) |
710 |
temperature |
Oil Cooling: after cooler glycol pipe |
Omega SA2C-RTD-3-100-A-40 |
711 |
oil pressure |
Oil Cooling: high pressure pump output (front panel) |
Omega PX359-2KGI across R=126.6 to convert 4.011-20.036 mA output. Convertion is made in E1242 |
801 |
temperature |
AHU East Glycol Supply |
Omega RTD PT100 surface |
802 |
temperature |
AHU East Glycol Return |
Omega RTD PT100 surface |
803 |
temperature |
AHU East Air Out |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
804 |
temperature |
AHU West Glycol Supply |
Omega RTD PT100 surface |
805 |
temperature |
AHU West Glycol Return |
Omega RTD PT100 surface |
806 |
temperature |
AHU West Air Out |
RTD PT100 Omega PRTF-10-2-100-1/4-6-E |
901 |
temperature |
FCS to Chiller2 |
Omega SA1-RTD-B Surface RTD |
902 |
temperature |
FCS from Chiller2 |
Omega SA1-RTD-B SURFACE RTD |
903 |
temperature |
BCS to Chiller1 |
Omega SA1-RTD-B SURFACE RTD |
904 |
temperature |
BCS from Chiller1 |
Omega SA1-RTD-B SURFACE RTD |
905 |
voltage dc |
FCS to Chiller2 |
Onicon F-1200 Dual Turbine Flow meter with D-1201 Display Module |
906 |
voltage dc |
BCS to Chiller1 |
Onicon F-1200 Dual Turbine Flow meter with D-1221 Display Module |
907 |
voltage dc |
AHU East Glycol Supply |
Omega FP-5300 with Signal Cond. Omega iDRN-FP-FS in: 6-60Hz out:0-10v |
908 |
voltage dc |
AHU West Glycol Supply |
Omega FP-5300 with Signal Cond. Omega iDRN-FP-FS in: 6-60Hz out:0-10v |
909 |
voltage dc |
AHU East Valve Feedback |
Voltage feedback from Glycol 3way valve of AHU East |
910 |
voltage dc |
AHU West Valve Feedback |
Voltage feedback from Glycol 3way valve of AHU West |
911 |
voltage dc |
AHU East Fan Freq Feedback |
Voltage feedback from VFD converted to Hz |
912 |
voltage dc |
AHU West Fan Freq Feedback |
Voltage feedback from VFD |
1011 |
temperature |
Glycol supply line to C Floor coil. MZ window at DECal Room |
Omega Surface RTD PT100 |
1012 |
temperature |
Glycol return line from C Floor coil. MZ window at DECal room |
Omega Surface RTD PT100 |
1013 |
temperature |
Glycol supply line to M Floor. P floor hatch |
Omega Surface RTD PT100 |
1014 |
temperature |
Glycol return line to M Floor. P floor hatch |
Omega Surface RTD PT100 |