Date:
Sunday, June 27, 2021
Telescope:
CTIO 1.5-m
Submitter:
Rodrigo Hinojosa
Submitter Email:
Observer Support:
None
Night Assistant:
Rodrigo Hinojosa
Problems:
File 1141 without structure (like bias), it was a calibration lamp in fiber mode taken right after GJ628. The PI_LAMPS.vi showed the message "no response". I restarted the GUI and then tried to Turn ON the lamp, but I received the message: Lamp device is OFFLINE. I checked the cable connections in the box of the lamps, they seemed to be ok, then David Rojas and I turned OFF/ON the power of the lamp box and the problem was solved, several spectra are named "junk" because of this problem.
Files 1185 and 1192 without structure, they both are lamps taken after a science observation. I restarted the GUI and then I could get the calibration lamps.
Program:
Target of Oportunity:
No
Time Observed:
11.0
Time Engineering:
0.0
Time lost technical:
0.5
Time lost weather:
0.0
Time lost other:
0.0
Primary Investigator:
Todd Henry
Email:
Institution:
SMARTS
Instrument:
CTIO 1.5m Echelle
Organization:
SMARTS
Total Program Time:
11.5
| Total Times | Time Observed | Time Engineering | Time lost technical | Time lost weather | Time lost other | Total Program Time |
|---|---|---|---|---|---|---|
| 2021-06-27 | 11.0 | 0.0 | 0.5 | 0.0 | 0.0 | 11.5 |
No time for HIP003535.
No weather data during the whole night.